|
IV. SAMPLE HANDLING

A. General Considerations

The only sample preparation necessary with FTIR-PAS
measurements
is to have a sample size smaller than 10 mm in diameter in order
to fit into the sample holder. The thickness should not exceed
6 mm. Any sample geometry can be directly analyzed including
disks, slabs, chips, chunks, pellets, powders, fibers, and sheets.
Semi-solid and liquids can also be analyzed. Larger samples are
cut or abraded to suitable dimensions, typically using razor
a blade, scissors, hacksaw, corker borer (#6), or a file or abrasive.
Figure 4 shows a typical sample ready for analysis.
Fig. 4 Sample holder with a polymer chunk sample
ready to
be inserted into the photoacoustic detector.

B. Sample loading

A MTEC sample holder is shown in Fig. 5 with large and small
stainless steel sample cups and brass spacers. The latter are
used to minimize the gas volume and thereby increase the photoacoustic
signal. A small cup is shown in Fig. 6 placed in the sample holder.
A funnel for loading powdered samples and other components for
handling samples are shown in the figure also. The sample should
be approximately one millimeter below the rim of the sample holder
in order to allow a layer of gas between the sample and the detector
window for photoacoustic signal generation in the gas. If very
high or low modulation frequencies are used, an approximate distance
that the sample should be below the cup rim can be calculated
from the formula where D =
1.51 cm2/s for helium and D= 0.187 cm2
for air. The value used for f, when a rapid scan FTIR spectrometer
is used, should be calculated from
were is the lowest wave number
value in the spectrum. |