|
Many samples have moisture contamination which will
cause the presence of water vapor in the detector's sample chamber
and the appearance of water vapor absorbance bands in spectra.
A desiccant for water vapor reduction can be put into a large
stainless steel cup as shown in Fig. 7 which is placed in the
sample holder underneath the sample cup. Fig. 8 shows a large
cup with polymer pellets which is placed above the desiccant
cup. As mentioned in Section III. A., magnesium perchlorate is
a very effective desiccant. Water vapor can be eliminated with
the combination of purging the detector with dry helium, allowing
adequate time for the desiccant to work, and minimizing the amount
of sample in the cup. Use only enough material to cover the bottom
of a small cup if water vapor is a problem. When not in use,
desiccant should never be left in cups open to the room's atmosphere
because moisture from the room will collect in the desiccant
and form a corrosive liquid.
Fig. 7. Background: Large stainless steel cup
loaded with
desiccant resting on a cup fixture. Foreground: Desiccant cup
has been placed underneath a slotted brass spacer insert in the
sample holder. Sample cups are placed on the spacer. See Fig.
8. |