|
has been plotted as a transmission spectrum using
the procedure
discussed in Section V.A.1. Many of the small-and medium-sized
features of the conventional transmission spectrum are so seriously
distorted that they are not discernible for a qualitative analysis
and that observable features are not useful for quantitative
analyses. For instance, it is not possible to observe the small
band at 2330 cm-1 in the transmission spectrum
and the larger band at 1100 cm-1 is not suitable
for quantitative analysis due to the fringe bands.
Fig. 33. Conventional transmission and FTIR-PAS spectra of a
32 micrometer thick polyethylene film. The PAS spectrum has been
converted to transmittance for the purpose of comparison. Note
the absence of interference fringes in the PAS spectrum.

2.
Polarized beam measurements on oriented films

In
production, polymer fibers are oriented by drawing in order to
improve tensile strength. Polarized FTIR-PAS measurements are
useful in monitoring the orientation process. A polarizer is
placed in front of the photoacoustic detector so that the infrared
beam incident on the sample in the detector's sample holder is
polarized. It may be useful to initially rotate the polarizer
until the signal is maximized with a carbon black reference in
the detector. This procedure will ensure that the polarizations
imposed by the FTIR optics and the polarizer are in coincidence
resulting in maximal intensity in the polarized beam.
After
this adjustment, the polarizer should be left fixed and the sample
should be rotated in the sample cup. Figure 34 shows spectra
of a PET film with the infrared beam polarized parallel and perpendicular
to the direction of draw. A number of spectral changes are observed
as a function of orientation particularly in the 800-900 cm-1
and 1300-1400 cm-1 ranges. The absence of
interference fringes in FTIR-PAS spectra is obviously also an
advantage in this application.
|