|
|
![]() |
![]() |
|
|
C. System Checks At initial set-up the optical alignment
of the photoacoustic
detector to the FTIR infrared beam should be checked and adjusted
if necessary. A liquid crystal infrared imager is supplied by
MTEC for this purpose. The imager is placed in the sample holder
and provides a clear direct image of the infrared beam's focal
spot location. The detector should be aligned in the FTIR sample
compartment by centering the focal spot in the sample holder.
Kinematic detector mount positioners are supplied by MTEC and
should be secured in place to register the detector alignment
position so that future re-installation does not require alignment.
Note that if the FTIR spectrometer in use has provision for automatic
alignment of the interferometer, this alignment routine should
be done prior to placing the photoacoustic detector in the FTIR.
The spectrometer's own detector should be used because it has
the correct position and active area size for the alignment procedure. Resolution = 8 cm-1 Source aperture = maximum Spectral Range = 400-4000 cm-1 Number of scans = 8 Apodization = Medium Beer-Norton Sample = MTEC carbon black reference Detector gas atmosphere = helium Some FTIR systems
may not allow
exact duplication of these parameters and may not perform well
enough to fully meet this noise specification. Others will exceed
this specification. The adjustments that will most likely improve
FTIR-PAS system performance (when necessary) are alignment of
the FTIR beam splitter, and tune-up of the FTIR's mirror velocity
stabilizing servo-loop at low velocities. |
|
|
![]() |